28 PKG08761-UMSR160
ENGLISH
INSTRUMENT PERFORMANCE
GAGE Resolution 0.4µin (0.01µm)
MEASUREMENT
Range (Ra) Up to 1600µin (40µm)
Range (Rz, Rv, Rp, Rt) Up to 7800µin (199µm)
Repeatability 2% of value + noise
Accuracy 5% of reading + 4µin (0.1µm)
Noise 4µin (0.1µm)
CALIBRATION
Process Automatic software calibration
Standard Able to calculate to ISO 4287 Roughness Standards
PARAMETERS
Standards ISO 4287
ISO 4287 (Roughness) Ra, Rz, Rp, Rv, Rt, Rz1max, Rsk, Rq, Rku
ISO (Primary) Pa, Pz, Pp, Pv, Pt
TECHNICAL
DATA OUTPUT On-Screen
Up to 5 results per page, selectable on-
screen graph
BATTERY
Charger Mini USB 5V 1A 110-240VAC 50/60 Hz
Charging Time 4 hours
POWER
Battery Life >10,000 measurements per charge
Standby Time 5,000 hours
Instant On Max 5 sec from standby to ready to measure
Auto-Sleep Function 5 minutes
INSTRUMENT CAPABILITY
PICK-UP ASSEMBLY
Pick-Up Type Piezoelectric
Stylus Type Diamond, Radius 200µin (5µm)
GAGE
Gage Force 200mg
Measurement Type Skidded
FILTER
Filter Type Gaussian
Filter Cut-Off 0.8mm
TRAVERSE
Traverse Length 0.2in (5mm)
Traverse Speed 0.08in/sec (2mm/sec)
DISPLAY Units µin/µm
ENVIRONMENTAL/PHYSICAL
PHYSICAL SPECIFICATIONS
Weight Including Pick-Up 14oz (0.4kg)
Power Source Li-Poly rechargeable battery
OPERATING CONDITIONS
Temperature 41 - 104°F (5 -40°C)
Humidity 0 - 80% non-condensing
STORAGE
Temperature 32 - 122°F (0 - 50°C)
Humidity 0 - 80% non-condensing
www.GlobalTestSupply.com