Instructions for use
Warmer for Blood, Intravenous
Fluids and Irrigation Fluids
REF AFP300
REF AFP302
STIHLER ELECTRONIC GmbH • 70597 Stuttgart • Germany
Part No. 10038.12 Rev. 05 10/2017

Table of Contents

Questions and Answers

  • E
    Elizabeth MurrayJul 25, 2025
    What to do if STIHLER ELECTRONIC ASTOFLO PLUS ECO shows overheating alarm?
    • S
      smithjakeJul 25, 2025
      If you're getting an overheating alarm on your STIHLER ELECTRONIC Medical Equipment, it could be due to an external heat source like sunlight or a radiator. Try to eliminate the heat source or move the equipment to a cooler location. It could also be that the ambient temperature is too high, so again, try to cool the environment. In some cases, the control unit or heating profile might be defective, and in that case, you should return it to your local sales office.
  • I
    imartinJul 30, 2025
    Why STIHLER ELECTRONIC ASTOFLO PLUS ECO Medical Equipment fails to enter standby mode?
    • L
      lambertdavidJul 30, 2025
      If your STIHLER ELECTRONIC Medical Equipment fails to enter standby mode, there might be a power supply issue or no power at all. Check the plugs and fuse, and make sure the mains voltage matches what's on the rating plate. Another possibility is that the control unit isn't properly plugged in. Try plugging it into a functioning socket, pressing the Standby button, and then pressing the Start button to switch to Heating Mode. If none of these steps work, the control unit may be defective, in which case you should return it to your local sales office.
  • C
    Carol GarciaAug 3, 2025
    How to fix low temperature alarm on STIHLER ELECTRONIC Medical Equipment?
    • K
      Kyle PhillipsAug 3, 2025
      If your STIHLER ELECTRONIC Medical Equipment is showing a low temperature alarm, it could be because the ambient temperature is too low. Try moving it to a warmer location. Another cause might be a defective heating profile, in which case you should return it to your local sales office. The connection cable for the heating profile could also be defective, requiring you to return the control unit to your local sales office.
  • M
    Matthew AustinAug 7, 2025
    What to do if STIHLER ELECTRONIC ASTOFLO PLUS ECO shows connection alarm?
    • D
      David CookAug 7, 2025
      If you receive a connection alarm on your STIHLER ELECTRONIC Medical Equipment, it means either no heating profile is connected, or it has become disconnected. To resolve this: 1. Connect the heating profile to the control unit. 2. Press the “Start” button to switch the device to Heating Mode.
  • K
    Kimberly MillerAug 11, 2025
    Why STIHLER ELECTRONIC ASTOFLO PLUS ECO shows self-test alarm?
    • N
      nelsonchristopherAug 12, 2025
      If your STIHLER ELECTRONIC Medical Equipment displays a self-test alarm, it might be because the Start button was pressed too soon. Clear the alarm and repeat the procedure. If the problem persists, it could indicate a failure in the control unit's electronics, requiring you to return the Warmer to your local sales office.
  • S
    Sandra MitchellAug 16, 2025
    What causes cable break alarm on STIHLER ELECTRONIC ASTOFLO PLUS ECO Medical Equipment?
    • S
      Sandra WilcoxAug 16, 2025
      If you're getting a cable break alarm on your STIHLER ELECTRONIC Medical Equipment, it suggests that either the control unit or the heating profile is defective. In this case, return the Warmer and/or heating profile to your local sales office.
  • J
    Joanne TateAug 20, 2025
    What to do if STIHLER ELECTRONIC ASTOFLO PLUS ECO Medical Equipment fails to turn on?
    • J
      Julie SmithAug 20, 2025
      If your STIHLER ELECTRONIC Medical Equipment fails to turn on, it means that the control unit is defective. Return the control unit to your local sales office.

Summary

3 Important Safety Information

Safety Levels and Definitions

Explains DANGER, WARNING, CAUTION, NOTICE levels and associated risks.

4 Specification of Application

6 Product Description

6.3 Components of the ASTOFLO PLUS ECO

Identifies and describes the physical parts of the ASTOFLO PLUS ECO.

6.4 Control Panel Explained

Detailed explanation of the control panel buttons, LEDs, and display.

8 Installation

8.2 Mounting the Warmer

Steps for attaching the device to infusion stands or medical rails.

9 Getting Started

9.2 Infusion Line Setup

Instructions for priming, inserting, and starting the infusion line.

10 Alarms and Troubleshooting

10.1 Low Temperature Alarm

Details the low temperature alarm, its causes, and actions.

10.2 Overheating Alarm

Explains the overheating alarm, its triggers, and resolution steps.

10.3 Cable Break Alarm

Describes the alarm for a broken cable or sensor and troubleshooting.

11 Brief Overview of Operating States and Alarms

12 Maintenance

12.1 Recurrent Testing

Outlines periodic tests required to ensure safe operation and performance.

Electrical Safety Test Setup

Details the setup for electrical safety tests including visual inspection and leakage currents.

Heating Profile Inspection and Tests

Procedures for visual inspection and performance tests of the heating profile.

12.2 Set Up for Electrical Safety Tests

Illustrates the test setup for various electrical safety measurements.

13 Technical Data

15 Ordering Information and Accessories

STIHLER ELECTRONIC ASTOFLO PLUS ECO Specifications

General IconGeneral
BrandSTIHLER ELECTRONIC
ModelASTOFLO PLUS ECO
CategoryMedical Equipment
LanguageEnglish