Known Use Errors
It is important to acknowledge and be aware of potential hazards associated with high suction
devices prior to use. This information highlights known use errors that constitute improper and
unsafe use of a high suction device, such as the Neptune 3 Rover. Stryker has received incident
reports of serious injury and/or death associated with operating high suction devices.
Stryker has received reports that high suction devices have been connected directly to patients’
chest tubes, postoperatively, which resulted in deaths.
Stryker has received reports that high suction devices have been connected to patients’ closed
wound drains, postoperatively, which resulted in serious patient injuries.
Do not connect the Neptune 3 Rover directly to a patient’s chest tube.
Do not connect the Neptune 3 Rover to a closed wound drainage system.
For training information, visit www.neptunecustomercare.com.