MAN0001-DFU • Rev. A • 11/12 Wallach Surgical Devices
User Manual for LifeDop
®
L150 Series (Handheld Doppler without Display)
L250 Series (Handheld Doppler with Display)
L150, L150A, L150R, L250, L250AR, L250R,
MC150, MC150R, MC250, MC250R
REF

Questions and Answers

  • J
    Jonathan PerezJul 31, 2025
    What to do if there is radio frequency interference with Summit Doppler Medical Equipment?
    • M
      Margaret RichardsonAug 1, 2025
      If you experience radio frequency interference with your Summit Doppler Medical Equipment, try increasing the distance between the LifeDop and the source of the interference.
  • A
    andrea20Aug 5, 2025
    Why is the heart rate inaccurate on my Summit Doppler Medical Equipment?
    • J
      Jennifer MartinezAug 5, 2025
      If the heart rate displayed on your Summit Doppler Medical Equipment is inaccurate, relocate the probe for a better signal. For OB applications, make sure that maternal sounds are not mixing with fetal sounds. Also, manually count to ensure the rate is between 50 and 220 BPM.
  • P
    Paul AdamsAug 9, 2025
    Why doesn't my Summit Doppler Medical Equipment rechargeable unit hold a charge?
    • J
      Joshua HillAug 9, 2025
      If your Summit Doppler Medical Equipment rechargeable unit does not hold a charge, first verify that you are using the correct recharge adapter, specifically Wallach Surgical Devices products. If the batteries are old, refer to the Maintenance Section.
  • N
    Nicole MorganAug 13, 2025
    Why is the recharge indicator flashing on my Summit Doppler Medical Equipment?
    • R
      richardbooneAug 13, 2025
      If the recharge indicator is flashing on your Summit Doppler Medical Equipment, it means the recharge cycle is complete or the batteries didn’t require recharge.
  • D
    Debra VaughnAug 18, 2025
    How to improve poor sound quality on Summit Doppler Medical Equipment?
    • S
      Shannon PughAug 18, 2025
      If the sound quality of your Summit Doppler Medical Equipment is poor, relocate the probe for a better signal. Refer to the Signal Strength Indicator.
  • M
    Martha JacksonAug 22, 2025
    What to do if probe frequency does not match the connected probe on Summit Doppler Medical Equipment?
    • W
      William AllenAug 22, 2025
      If the probe frequency does not match the connected probe on your Summit Doppler Medical Equipment, check that the attached probe is the correct one and properly attached. If the correct probe is being used, contact Wallach Surgical Devices’ Service Department.
  • I
    Isaiah CaseyAug 27, 2025
    Why is the recharge indicator off after charging Summit Doppler Medical Equipment?
    • C
      christinafoxAug 27, 2025
      If the recharge indicator is off after charging your Summit Doppler Medical Equipment, a battery level error has occurred. Refer to Diagnostic Codes and contact service.

Summary

Intended Use

Obstetric Use Cases

Detects fetal heart beats for viability assessment using 2 and 3 MHz probes.

Vascular Use Cases

Detects blood flow in veins/arteries for peripheral vascular disease detection (4, 5, 8 MHz).

Device Warnings and Contraindications

Lists contraindications, warnings, and cautions for device operation.

Ultrasound Safety Guidelines

Ultrasound Probe Specifications

Obstetrical Probe Frequencies Explained

Details on 2 MHz (late term) and 3 MHz (general purpose) obstetrical probes.

Vascular Probe Frequencies Explained

Details on 4 MHz (broad), 5 MHz (deep vessels), and 8 MHz (shallow vessels) vascular probes.

Unit Operation & Display

Device Power On;Off Operation

Instructions for powering the unit on/off and its auto-shutdown feature.

Battery Management

Battery Level Monitoring

Visual indicators for monitoring battery status on display and non-display units.

Recharging the Device Battery

Steps and warnings for recharging the device's batteries.

PC Audio Download - Windows

PC Audio Download - Apple

Obtaining Doppler Signals

Coupling Gel Usage Importance

Importance of adequate gel for ultrasound transmission and probe performance.

Obstetrical Signal Acquisition Techniques

Techniques for locating fetal heart sounds at different stages of pregnancy.

Advanced Signal Acquisition

Vascular Signal Location Techniques

Guidance for obtaining peripheral arterial and venous sounds using 5 and 8 MHz probes.

Device Maintenance and Cleaning

Post-Examination Cleaning

Guidelines for cleaning the unit and probes after each examination.

Disinfection and Storage Guidelines

Recommended disinfectants and proper storage conditions.

Battery Replacement Guide

How to Replace Device Batteries

Step-by-step instructions for safely replacing batteries.

Troubleshooting Common Issues

Troubleshooting Sound and Rate Issues

Common causes and solutions for poor sound quality and inaccurate heart rate.

Interference and Diagnostics

Warranty and Service Information

Summit Doppler LifeDop L150 Series Specifications

General IconGeneral
CategoryMedical Equipment
TypeDoppler
ModelLifeDop L150 Series
DisplayLCD
Power SourceRechargeable battery or AC adapter
Probe TypeInterchangeable
ApplicationVascular
Frequency8 MHz
Battery LifeApproximately 8 hours
WeightApproximately 0.7 lbs (0.32 kg)
OutputAudio