7
MAINTENANCE
MAINTENANCE OF PROBE
If gaging action becomes sluggish or if gaging points stick open, probe
may be disassembled for cleaning (see Figure 7).
1. Loosen Set Screw with 5/64" (2 mm) Hex Key Wrench and remove
Head.
2. Remove Wedge from Tube carefully. Do not lose Spring.
3. Flush Tube with an oily solvent and then check point movement
using Wedge.
4. Relubricate Wedge with 5-10 SAE machine oil and reassemble. Be
certain Screw engages countersink in Tube.
5. Check Wedge movement. Spring must push Wedge out to release all
pressure on Gaging Points.
MAINTENANCE OF RETRACTING LEVER
If Retracting Lever becomes sluggish in operation, it may be removed
for cleaning (see Figure 8).
1. Insert .050" (1.25 mm) Hex Key Wrench into small hole in Lever.
Loosen approximately one turn.
2. Remove Wrench and use end to push Shaft out of body.
FIGURE 8
SHAFT
.050”
HEX KEY WRENCH
RETRACTING
LEVER