EasyManua.ls Logo

TechTest RECOGNOIL - Settings and Oil-Pick Dialog

TechTest RECOGNOIL
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Recognoil – User‘s Manual
Software version 2.2.1_2015-02-21
-
16
-
3.6. Settings and Oil-Pick dialog
img. 11. The Options dialog
Validation - calibration module: Please see chapter 6.
Working directory: click Change if you want to choose other directory, which
contains oil files, RAW files and Result files.
Manual exposition in case of need of very short or very long exposure times, use
this option. The Detector output outcome may not be the same as in the automatic
intelligent mode. For advanced users only.
You can disable the button sound effects by un-checking “Button sound effects”
Save RAW images – the default setting is ON – the software records every detection
into a RRW file, which can be later opened and evaluated. Every files has approx.
0.5 Mbytes.
Always use ALD option in automatic intelligent mode, the ALD method is
activated, when there is the environment light getting to the detector. If you want
to use this mode every detection, regardless on this amount of light, check this
option. The detection then takes twice the time