TROUBLE-SHOOTING GUIDE
Blower not ON
No airflow from blower
Problems Trouble-shootingCauses
Airflow test failed
Battery time is too short
even fully charged
Increased sound level
Warning indicator ON,
blower vibrate and alarm
sound bleeping
Feeling smell of incoming air
Long press ON button.
Charge the battery.
Check and reassemble the
battery.
Check and clear the
obstruction.
Check the tube status.
Replace new filter.
Replace new battery.
Fully charge battery.
Replace filter.
Replace a new charger.
Replace filter and pre-filter
as required.
Check if tube/anywhere gets
blocked before use.
Check if the package is
removed.
Check the filter status and
replace new one if needed.
Check how the tube
assemble as well as status.
Check and equip both filters.
Battery no power
Battery not installed properly
Tube blocked/air leakage
Dirty filter needs replacement
Battery faulty
Incorrect charging
Blocked filter
Damaged charger
Filter is getting clogged
Damaged Filter
Hose with leakage problem
Filter component not
complete
Tube gets blocked/air
leakage
Filter assembled without
removing the package
The hose may get blocked/
air leakage
STORAGE
The TECMEN PAPR system is not intrinsically safe. Keep away from flammable, or
explosive atmosphere. Storage should be in a clean, dry, cool place with filter.
Blower storage
Stored at a temperature between -4°F to 113°F (-20°C to 45°C), in a clean environment
without direct light.
Battery storage
To help maximize battery service life:
11
Disconnect the charger after a full charge has been received.
Battery should be removed from blower if long time storage.
Store the battery at 14°F to 115°F (-10°C to 45°C), R.H. <85%, to get maximize battery
service life.
•
•
•