PROVASET T3LPF REV.20210610
87
Press the message on the screen: "FOL-
LOW THE INSTRUCTIONS!"
CALIBRATION
PAY ATTENTION:
Editing these parameters could
affect the testing functioning!
Select "Relative pressure" or "Flow".
CALIBRATION
PRESSURE
FLOW
11.1 Calibration of the relative pressure
The relative pressure is the pressure in the component being tested, in relation to the
Calibration procedure:
A
B
TEST
LTC
+
1. Switch on the equipment and wait at least 10 minutes for the pressure read-
ings to stabilize;