0. COMMISSIONING AND PREPARING FOR STERILIZATION
1. Fill in the water tank (see paragraph 8);
2. Push the MAIN BUTTON to start the machine;
3. Open the door 1A;
4. Push the button PUMP to let the water in the chamber.
Photo E
Photo E
Autoclave Time (sec.)
HYDRA EVO 20 30
HYDRA EVO PLUS 25 35
ATTENTION: If you try to start the cycle up without respecting this minimum period the
autoclave is showing ERROR 400.
5. Leave the button PUMP to stop the water inlet.
ATTENTION: LET ALWAYS WATER INLET BEFORE STARTING THE STERILISATION
CYCLE UP!
6. Put the instruments on the trays according to the statements shown at the beginning of the
manual;
7. It is very important not to lay the instruments upon another. Do not overcharge the trays.
8. Each time you start a sterilization cycle up you have to put in the chamber an integrator to
receive a confirmation of the sterilization. We suggest to put the integrator in the middle of the
chamber.
9. Set the required sterilization cycle (Cycle 121°C or Cycle 134°C ) pressing the button SELECT.
10. Close and block the door and at the same time let the cycle start. If this is not happening in the
stabilized time the autoclave is showing the ERROR 450.
ATTENTION: LET THE CYCLE START UP AND CLOSE THE DOOR AT THE SAME
TIME
It is important that you
check continuously that
the level of the water is
not exceeding the metal
tablet so that the water
is not going out.