1. MAINTENANCE
CAUTION: DISCONNECT ALWAYS THE MAIN SUPPLY BEFORE EACH TECHNICAL
INTERVENTION!
11.1. DAILY ROUTINE MAINTENANCE
• Clean the seal on the door 1A and on the outer edge of the sterilisation chamber sealing
the gasket using a wet cloth or the soft part of the supplied sponge.
This cleaning operation must be carried out to remove possible dirt which can cause a
pressure loss in the sterilisation chamber and a possible seal cut.
• Check the water level in the tank (see paragraph 8).
11.2. WEEKLY ROUTINE MAINTENANCE
• To remove the deposits on the chamber floor, clean with the abrasive part of the supplied
sponge after removing the tray-holder.
To moisten the sponge, only use distilled or demineralized water.
• Empty the water tank (see paragraph.9).
11.3. QUARTERLY MAINTENANCE
• Lubricate the hinges and the closing pin with silicon oil.
• Adjusting the door 1A :
If the handle 2A is too frail while closing and blocking the door, turn it counterclockwise
Photo F to increase the pressure of the gasket on the chamber.