10.2
Errors
In case of error alarm the device has to be reset to start a new cycle.
ALARM RESET
Display
Cause
Solution
Water infeed alarm
- the tap is closed
- filters clogged
- pressure
of the water
system not normal
- Make sure the tap is open.
- Clean water infeed filter.
- Make sure water pressure ranges are
within the indicated values. If the pressure
is insufficient, contact a plumber.
- Make sure the Water Block (OPTIONAL)
is not locked.
- Make sure there is no water into the tank
and start a new cycle
Water level alarm
- the device has been
moved or tilted during the
cleaning.
- too many instruments have
been added after water
infeed.
- Make sure there is no water into the tank
and start a new cycle.
- Do not add instruments during the cycle.
Water drain alarm
- filters clogged
- Discharge pump breakage
- Clean the tank filter.
- Make sure there is no water into the tank
and start a new cycle.
Keep pressed for 8s. button
OK