Maintenance
and
Calibration
Type
114
f.
Using
Table
4-2,
check
the
period
accuracy
of
the
re
maining
settings
of
the
PERIOD
switch.
In
all
cases
the
ac
curacy
should
be
as
specified.
9.
Check
PERIOD
VARIABLE
control
for
a
range
of
10:1
or
more
a.
Set
the
Type
114
PERIOD
control
to
10
mS
and
the
WIDTH
control
to
100
µ
S.
b.
Turn
the
Type
1A1
MODE
switch
to
CH
1.
c.
Set
the
Type
547
TIME/CM
|B|
to
.1
SEC.
d.
Rotate
the
PERIOD
VARIABLE
control
clockwise.
Check
that
a
display
of
1
marker/cm
is
obtainable
(
10:
1).
10.
Adjust
WIDTH
CAL
control
R180
a.
Set
the
Type
114
PERIOD
and
WIDTH
controls
to
1
mS
and
turn
both
VARIABLE
controls
to
CAL
(see
Fig.
4-11).
b.
Set
the
Type
180
A
for
100
microsecond
markers.
C.
Turn
the
Type
1A1
MODE
switch
to
ALT.
d.
Set
the
Type
547
TIME/CM
(
B
)
to
.2
mSEC
and
adjust
the
TRIGGERING
LEVEL
control
for
a
stable
display.
e.
Use
the
Type
1A1
POSITION
controls
to
obtain
the
dis
play
shown
in
Fig.
4-12.
f
Adjust
the
WIDTH
CAL
control
R180
(see
Fig.
4-13)
for
a
positive
pulse
width
of
1
millisecond
(ten
100
microsecond
periods)
using
the
time
markers
from
the
Type
180A
as
reference.
g.
Check
to
see
that
WIDTH
>
PERIOD
light
is
on.
Fig.
4-11.
Control
settings
for
adjusting
WIDTH
CAL
control
R18O.
TABLE
4-2
Type
114
PERIOD
Type
114
WIDTH
Type
547
TIME/CM
(B)
Type
180A
Markers
Error
10
mS
10
µS
10
mSEC
10
milliseconds
±3%
1
mS
10
µ
S
1
mSEC
1
millisecond
±3%
100
µ
S
1
µ
S
.1
mSEC
100
microseconds
Adjusted
to
<
½%
10
µ
S
100
nS
10
µ
SEC
10
microseconds
±3%
1
µ
S
100
nS
1
µ
SEC
1
microsecond
Adjusted
to
<
3%
4-8