Maintenance
and
Calibration
Type
114
Fig,
4-12.
Crf
display
showing
correct
adjustment
of
WIDTH
CAL
control
R180.
11.
Adjust
100
nS
Pulse
Width
capacitor
C195F
©
a.
Set
the
Type
114
WIDTH
switch
t
o
100
nS,
set
the
PE
RIOD
switch
to
1
µ
S
(see
Fig.
414).
b.
Set
the
Type
547
TIME/CM
(B)
to
.1
µ
SEC.
c.
Set
the
Type
1A1
Channel
1
VOLTS/CM
switch
t
o
2.
Rotate
the
Type
114
AMPLITUDE
VARIABLE
and
the
Type
1A1
Channel
1
POSITION
controls
lor
5
cm
of
pulse
ampli
tude.
Center
the
pulse
vertically
on
the
graticule.
Set
the
Type
547
TRIGGERING
LEVEL
control
for
stable
triggering.
d.
Set
the
Type
180A
to
deliver
10
m
c
sine-wave
output.
e.
Use
the
Type
1A1
Channel
2
POSITION
control
to
posi
tion
the
top
of
the
sine
waves
from
the
Type
180A
to
the
graticule
centerline.
Fig.
4-13.
location
of
WIDTH
CAL
control
R180
f.
Set
the
SWEEP
MAGNIFIER
control
to
X5.
Use
the
HORI
ZONTAL
POSITION
and
TRIGGERING
LEVEL
controls
to
obtain
the
display
shown
in
Fig.
4-15.
NOTE
At
100
nanosecond
pulse
widths,
the
risetime
and
falltime
of
the
pulse
becomes
a
significant
portion
of
the
pulse
width,
therefore
the
pulse
width
meas
urement
is
made
at
the
50%
amplitude
points.
Fig.
4-14.
Control
sortings
for
adjusting
l
00
nS
Pulse
Width
capacitor
C195F.
4-9