Maintenance
and
Calibration
Type
114
g.
Ad|ust
C195F
(see
Fig.
4-16)
for
100
nanoseconds
pulse
width.
Allowable
error
is
±
3%,
however,
it
is
possible
to
adjust
more
closely
than
this.
Fig.
4-15.
Crt
display
for
proper
setting
of
100
nS
Pulse
Width
capacitor
C195F.
14.
Check
operation
in
square-wave
mode
a.
Set
the
Type
114
WIDTH
control
to
SQUARE
WAVE.
b.
Using
Table
4-4
check
square
wave
operation
for
all
positions
of
the
PERIOD
control.
Allowable
period
error
is
4
mm
within
the
center
8
cm
of
the
trace
TABLE
4-4
Type
114
PERIOD
Type
547
TIME/CM
(B)
Cycles/8
cm
10
mS
10
mSEC
8
1
mS
1
mSEC
8
100
µ
S
.1
mSEC
8
10
µ
S
10 µ
SEC
8
1
µ
S
1
µ
SEC
8
12.
Check
pulse
width
accuracy
Using
the
settings
given
in
Table
4-3,
check
the
pulse
width
accuracy
in
the
other
settings
of
the
WIDTH
control.
TABLE
4-3
Type
114
PERIOD
Type
114
WIDTH
Type
547
TIME/CM
Type
180A
Time
Markers
Number
of
Markers
1
mS
1
mS
.2
mSEC
100
µ
sec
10
1
mS
100
µ
S
20
µ
SEC
10
µsec
10
100
µ
S
10
µ
S
2
µ
SEC
1
µsec
10
10
µ
S
1
µ
S
.2
µ
SEC
10
m
c
10
cycles
13.
Check
WIDTH
VARIABLE
control
range
a.
Set
the
Type
114
WIDTH
control
to
100^5,
set
the
PE
RIOD
control
to
10
mS
and
the
PERIOD
VARIABLE
control
to
CAL.
b.
Set
the
Type
180A
for
100
microsecond
markers.
c.
Set
the
Type
1A1
Channel
1
VOLTS/CM
switch
to
5.
d.
Set
the
Type
547
TIME/CM
(B)
to
.2
mSEC,
the
TRIG
GERING
SLOPE
switch
to
+,
and
adjust
the
TRIGGERING
LEVEL
control
for
a
stable
display.
e.
Slowly
rotate
the
WIDTH
VARIABLE
control
clock
wise.
Check
for
a
pulse
width
of
at
least
1
millisecond
(ten
100
microsecond
markers)
at
the
clockwise
end
of
rotation
(
see
Fig.
4-12).
f.
Return
the
WIDTH
VARIABLE
control
to
the
CAL
position.
15.
Check
output
amplitude
(into
the
5012
ter
mination)
a
.
Turn
the
Type
114
AMPLITUDE
VARIABLE
control
slowly
counterclockwise,
checking
for
smooth
mechanical
and
electrical
oporation.
b.
Set
the
Type
1A1
Channel
1
VOLTS/CM
switch
to
1
and
check
for
3
cm
or
less
of
vertical
deflection
when
the
AMPLITUDE
VARIABLE
control
is
fully
counterclockwise.
Fig.
4-16.
location
of
100 nS
Pulse
Width
capacitor
C195F.
c.
Set
the
Type
1A1
Channel
1
VOLTS/CM
switch
to
2.
Rotate
the
Type
114
AMPLITUDE
VARIABLE
control
fully
clockwise.
Check
for
5
cm
or
more
of
vertical
deflection.
d.
Change
the
AMPLITUDE
switch
to
3
to
10
V
and
re
peat
the
amplitude
check.
Check
both
the
+
and
1
V
to
3
V
positions
of
the
AMPLITUDE
switch.
4-10