Performance Check/Calibration—Type 422 AC-DC
TABLE 5-1
Power Supply Output Voltage and Ripple
Power Supply
Output Voltage Range
Maximum Ripple2 (peak to peak)
High Frequency
AC Line Frequency
12 volt
11.9 to 12,1 volts
20 millivolts
8 millivolts
+ 12 volt
+ 11.7 to + 12.3 volts
20 millivolts 8 millivolts
+ 55 volt
+ 53.9 to 50.3 volts
2 volt
+ 95 volt
+ 93.1 to + 97 9 volts
0.4 volt
+ 10.5 volt
+ 10.3 to + 10.7 volts
1 millivolt 1 m illivolt
-81 volt
80 to 82 volts
20 millivolts 8 millivolts
2Disregard fast spikes; see Fig. 5-5.
Fig. 5-4. Location of +10.5-volt and -81-volt power supply test
points (Calibrator and Regulators board).
f. CHECKEach supply with the precision DC voltmeter
for the tolerances given in Table 5-1.
g. Disconnect the voltmeter.
4. Check DC Ripple
a. Set the test oscilloscope for a vertical deflection
factor of 10 millivolts/division, AC coupled, at a sweep rate
of 20 microseconds/division (IX probe connected to
vertical input connector).
b. CHECKTest oscilloscope display for maximum
ripple of each supply as listed in Table 5-1. Fig. 5-5 shows a
typical display of ripple and the levels between which the
ripple measurement should be made. Power supply test
points are shown in Fig. 5-2 and 5-4. Change the test
oscilloscope vertical deflection factor as necessary to
provide the correct display of ripple.
c. Set the variable DC power supply for +22-volts
output.
d. CHECKRipple of each supply for tolerances given in
Table 5-1.
c. Set the variable DC power supply for +22-volts out
put.
d. CHECKEach supply with the precision DC volt
meter for the tolerances given in Table 5-1.
e. Set the variable DC power supply for +35-volts
output.
e. Set the variable DC power supply for +11.5-volts out
put.
f. CHECK-Ripple of each supply for tolerances given in
Table 5-1.
g. Disconnect the variable DC power supply and the test
oscilloscope.
5-11