Performance verification
Harmonic Distortion Test Record (cont.)
Harmonic Distortion
(amplitude 1 V
p-p
)
Fundamental
= reference 2nd 3rd 4th 5th Limit
CH2 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-60 dBc
Sine 100 kHz
100 kHz
200 kHz 300 kHz
400 kHz 500 kHz
CH1 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-60 dBc
CH2 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-60 dBc
Sine 1 MHz
1 MHz 2 MHz 3 MHz 4 MHz 5 MHz
CH1 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-50 dBc
CH2 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-50 dBc
Sine 25MHz
25 MHz 50 MHz 75 MHz 100 MHz 125 MHz
CH1 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-40 dBc
CH2 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-40 dBc
AFG305xC Spectrum Analyzer reading
Sine 20 kHz
20 kHz 40 kHz 60 kHz 80 kHz 100 kHz
38 AFG3000C and AFG3000 Series Specifications and Performance Verification