EasyManua.ls Logo

Tektronix CFG253

Tektronix CFG253
19 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CFG253 Performance Verification & Adjustment Procedures
Page 5 of 19 Procedure #: CP1005 Revision: A
l. Set the FUNCTION to Square Wave. Use the CFG253 and test scope settings in
Table 1 to check the MAIN OUTPUT AMPLITUDE for the indicated settings :
TABLE 1
Function Range Volts Out
button
Freq.
Dial
Test Scope
Volts/Div
Test Scope
Display
Square Wave 1K Out 3.0 2V/Div > 5 Div
Sine Wave 1K Out 3.0 2V/Div > 5 Div
Sine Wave 1K Out 0.3 2V/Div > 5 Div
Square Wave 1K Out 0.3 2V/Div > 5 Div
Triangle 1K Out 0.3 2V/Div > 5 Div
m. Remove the 50 ohm terminator and connect the coax cable directly to the test
scope input.
n. Set the test scope Time/Div to 1uSec.
o. Check the test scope display for each setting in Table 2:
TABLE 2
Function Range Volts
Out
Freq.
Dial
Volts/D
iv
Display
Triangle 1M Out 3.0 5V/Div > 4 Div
SquareWave 1M Out 3.0 5V/Div > 4 Div
Sine Wave 1M Out 3.0 5V/Div > 4 Div
p. Press the VOLTS OUT button in. Check the test scope display for the settings in
Table 3:
TABLE 3
Function Range Volts
Out
Freq.
Dial
Volts/Div Display
Sine Wave 1M In 3.0 0.5V/Div > 4 Div
Square Wave 1M In 3.0 0.5V/Div > 4 Div
Triangle 1M In 3.0 0.5V/Div > 4 Div

Other manuals for Tektronix CFG253

Related product manuals