Commands Listed in Alphabetical Order
single sequenc
e acquisition will be started. If the last acquisition was continuous,
a new continuous acquisition will be started.
If RUN is issue
d in the middle of completing a single sequence acquisition (for
example, averaging or enveloping), the acquisition sequence is restarted, and
any a ccumulated data is discarded. Also, the o scilloscope resets the number of
acquisitions. If the RUN argument is issued while in continuous mode, acquisition
continues.
Group
Acquisition
Syntax
ACQuire:STATE {OFF|ON|RUN|STOP|<NR1>}
ACQuire:STATE?
Related Commands
ACQuire:STOPAfter
Arguments
OFF stops acquisitions.
STOP stops acquisitions.
ON starts acquisitions.
RUN starts acquisitions.
<NR1> = 0 stops acquisitions; any other value starts acquisitions.
Examples
ACQUIRE:STATE RUN starts the acquisition of waveform data.
ACQUIRE:STATE? might return:ACQUIRE:STATE 0 indicating that the
acquisition is stopped.
2-62 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2014B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2014B and is the answer not in the manual?

Tektronix DPO2014B Specifications

General IconGeneral
BrandTektronix
ModelDPO2014B
CategoryTest Equipment
LanguageEnglish

Related product manuals