2260B Series Programming Manual
16
4. Use the Current knob to set the
parameter for the chosen F setting.
5. Press the Voltage knob to save the
configuration setting. Conf will be
displayed when successful.
Exit
Press the Function key again to exit
the configuration settings. The
function key light will turn off.
Configuration Table
Please use the configuration settings listed below when applying
the configuration settings.
Normal Function
Settings Setting Setting Range
Output ON delay time F-01 0.00s~99.99s
Output OFF delay time F-02 0.00s~99.99s
V-I mode slew rate select
F-03
0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority
Rising voltage slew rate F-04
0.01V/s~60.00V/s (2260B-30-XX)
0.1V/s~160.0V/s (2260B-80-XX)
Falling voltage slew rate F-05
0.01V/s~60.00V/s (2260B-30-XX)
0.1V/s~160.0V/s (2260B-80-XX)
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2260B Series and is the answer not in the manual?

Tektronix Keithley 2260B Series Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2260B Series
CategoryPower Supply
LanguageEnglish

Related product manuals