REMOTE CONTROL
25
F-46 = 0 Subnet Mask part 4 of 4
F-43 = 172 Gateway part 1 of 4
F-44 = 16 Gateway part 2 of 4
F-45 = 21 Gateway part 3 of 4
F-46 = 101 Gateway part 4 of 4
F-57 = 1 Enable Sockets
Note
The socket function is only available for firmware
version V1.12 or above. See the user manual to
check your firmware version number.
USB Remote Control Function Check
Functionality
check
Invoke a terminal application such as Hyper
Terminal.
To check the COM port No, see the Device
Manager in the PC. For WinXP; Control panel
System Hardware tab.
Run this query command via the terminal after
the instrument has been configured for USB
remote control (page 21).
*idn?
This should return the Manufacturer, Model
number, Serial number, and Firmware version
in the following format.
XXXXXX,AAAAAA,TW123456,01.00.20110101
Manufacturer: XXXXXX
Model number : AAAAAA
Serial number : TW123456
Firmware version : 01.00.20110101
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2260B Series and is the answer not in the manual?

Tektronix Keithley 2260B Series Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2260B Series
CategoryPower Supply
LanguageEnglish

Related product manuals