EasyManuals Logo
Home>Tektronix>Test Equipment>MSO2012B

Tektronix MSO2012B Programmer's Manual

Tektronix MSO2012B
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #223 background imageLoading...
Page #223 background image
Commands Listed in Alphabetical Order
MEASUrement:
METHod
Sets or returns the method used to calculate the 0% and 1 00% reference level.
Group
Measurement
Syntax
MEASUrement:METHod {Auto|HIStogram| MINMax}
MEASUrement:METHod?
Related Commands
MEASUrement:REFLevel:PERCent:HIGH, MEASUrement:REFLevel:PERCent:
LOW, MEASUrement:REFLevel:PERCent:MID, MEASUrement:REFLevel:
PERCent:MID2
Arguments
Auto selects the best method for each data set.
HIStogram sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax uses the highest and lowest values of the waveform record. This selection
is best for examining waveforms with no large, flat portions of a common value,
such as sine waves and triangle waves.
Examples
MEASUREMENT:METHOD? might return :MEASUR EMENT:METHOD MINMAX
indicating that the reference levels are set to MIN and MAX.
MEASUrement:REFLevel? (Query Only)
Returns the current reference level parameters.
Group
Measurement
Syntax
MEASUrement:REFLevel?
Examples
MEASUREMENT:REFLEVEL? might return these reference
level settings
:MEASUREMENT:REFLEVEL:METHOD
PERCENT;ABSOLUTE:HIGH 0.0000;LOW 0.0000;MID1 0.0000 ;MID 2
0.0000;:MEASUREMENT:REFLEV EL:PERCENT:HIGH 90.0000;LOW
10.0000;MID1 50.0000 ;MID2 50.0000
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-207

Other manuals for Tektronix MSO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2012B and is the answer not in the manual?

Tektronix MSO2012B Specifications

General IconGeneral
BrandTektronix
ModelMSO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals