Commands Listed in Alphabetical Order
*CAL? (Query O
nly)
Performs an internal self-calibration and returns the oscilloscope calibration status.
NOTE. Disconnect or otherwise remove all input signals prior to starting
self-calib
ration. The self-calibration can take several minutes to complete.
No other commands are executed until calibration is complete.
Group
Calibration and Diagnostic
Syntax
*CAL?
Returns
<NR1> = 1 indicates the calibration did not complete successfully.
<NR1> = 0 indicates the calibration completed without errors.
Examples
*CAL? starts the internal signal path calibration and might return 0 to indicate that
the calibration was successful.
CALibrate:FACtory (No Query Form)
Provides the controls for starting and stopping the factory calibration process. The
factory calibration process consists of a series of steps.
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory {STARt|CONTinue|P REVious|ABOrt|DU mp}
Arguments
STARt initializes the factory calibration sequence and starts the first c alibration
step.
CONTinue begins the next factory calibration step.
PREVious attempts to run the most recent factory calibration step again.
ABOrt stops the calibration process.
DUmp stops the calibration and prints the calibration constants.
Examples
CALIBRATE:FACTORY START starts the calibration p rocess.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-95