Commands Listed in Alphabetical Order
Group
Waveform Trans
fer
Syntax
DATa:SOUrce {CH1|CH2|CH3|CH4|MATH|REF1 |REF2|
D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D12|D13|D14|D15}
DATa:SOUrce?
Related Commands
CURVe
Arguments
CH1–CH4 species which analog channel data will be transferred from the
oscilloscope to the contro ller, channels 1 through 4.
MATH species that the Math w aveform data will be transferred from the
oscilloscope to the controller.
REF1–REF2 species which Reference waveform data will be transferred from
the oscilloscope to the controller, waveforms, 1 through 4.
D0–D15 species that the returned waveform data is packaged as 4 bytes per
sample, representing the states of all digital channels (D0–D15) and the logic
states of the analog channels (Ch1–Ch4). When the
:DATA:SOURCE is set to
DIGITAL, the
:DATA:WIDTH is automatically set to 4.
The format of the returned 4- byte (32 bit) data is as follows:
2-130 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual