Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:I2C:DATa:VALue
Sets or returns the binary data string used for I2C triggering if the trigger condition
is DATA or ADDRANDDATA. Applies to bus <x>, where x is the bus number.
Conditions
This command requires a DPO2EMBD application module.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:I2C:DA Ta:VALue <QString>
TRIGger:A:BUS:B<x>:I2C:DA Ta:VALue?
Arguments
<QString> is the binary data string, where the number of bits is 8 times the
number of bytes specified. The only allowed characters in the string are 0, 1,
and X.
TRIGger:A:BUS:B<x>:LIN:CONDition
Sets or returns the trigger condition for LIN.
Conditions
This command requires a DPO2AUTO application m odule.
Group
Bus
Syntax
TRIGger:A:BUS:B<x>:LIN:CO NDition
{SYNCField|IDentifier|DAT A|IDANDDATA|WAKEup|SLEEP|ERROR}
TRIGger:A:BUS:B<x>:LIN:CO NDition?
Arguments
SYNCField sets the LIN trigger condition to sync field.
IDentifier sets the LIN trigger condition to identifier.
DATA sets the LIN trigger condition to data.
IDANDDATA sets the LIN trigger condition to ID and data.
WAKEup sets the LIN trigger condition to wake up.
SLEEP sets the LIN trigger condition to sleep.
ERROR sets the LIN trigger condition to error.
2-304 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual