EasyManua.ls Logo

Tektronix MSO2024B

Tektronix MSO2024B
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands Listed in Alphabetical Order
Examples
TRIGGER:EXTER
NAL:PROBE?
might return :TRIGGER:EXTERNAL:PROBE
1.0E1
for a 10X probe.
TRIGger:EXTernal:YUNIts? (Query Only)
Returns the external trigger vertical (Y) units value.
Group
Trigger
Syntax
TRIGger:EXTernal:YUNIts?
Examples
TRIGGER:EXTERNAL:YUNITS? might return TRIG GER:EXTERNAL:YUNITS
“V”
if the vertical unit is volts.
TRIGger:FREQuency? (Query Only)
Returns the available trigger frequency in hertz. If the trigger frequency is not
currently available, the IEEE Not A Number (NaN = 9.9100E+37) value is
returned. The maximum precision of the r eturned frequency is 6 digits.
Group
Trigger
Syntax
TRIGger:FREQuency?
TRIGger:STATE? (Query Only)
Returns the current state of the triggering system: A RMED, AUTO, READY,
SAVE
or TRIGGER.
Group
Trigger
Syntax
TRIGger:STATE?
Related Commands
TRIGger:A:MODe
Returns
ARMED indicates that the oscilloscope is acquiring pretrigger information.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-347