Reference Notes
Procedure B : Maximum
Allowable Pulse Amplitude
To determine th
e maximum allowable pulse amplitude do the following:
1. Measure the pulse width at the 50% points.
2. Divide the Ampere-second (or Ampere-microsecond) specification for the
range setting of the probe by the pulse width. The quotient is the maximum
allowable p
ulse amplitude; the peak amplitude of the measured pulse must
be less than this value.
For example
, the TCP312A Current Probe has a maximum Ampere-second
product of 500 A-μs in the 10 A/V range setting. If a pulse measured with a
TCP312A probe has a width of 15 μs, the maximum allowable peak current
would be 500 A-μs divided by 15 μs, or 33.3 A.
TCPA300/400 Amplifiers and TCP300A/400 Series Current Probes User Manual 29