Performance Tests
TDS Family Performance Verification and Specifications
1–87
4. Set the generator to the new frequency and reference level:
H Change the sine wave generator to the desired new frequency.
H Input the correction factor for the new frequency into the level meter.
H Adjust the sine wave generator amplitude until the level meter again
reads the value noted in step 3. The signal amplitude is now correctly set
for the new frequency.
H Disconnect the sine wave generator from the power sensor.
H Connect the sine wave generator to the digitizing oscilloscope.