EasyManua.ls Logo

Tektronix TDS 784D - Page 321

Tektronix TDS 784D
376 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B: Algorithms
TDS 500D, TDS 600B, & TDS 700D User Manual
B–11
Timing measurement. The ratio of the negative pulse width to the signal period
expressed as a percentage.
NegativeWidth is defined in Negative Width, below.
If Period = 0 or undefined then return an error.
NegativeDutyCycle =
NegativeWidth
Period
100%
Amplitude (voltage) measurement.
NegativeOvershoot =
Low * Min
Amplitude
100%
Note that this value should never be negative (unless High or Low are set
out-of-range).
Timing measurement. The distance (time) between MidRef (default = 50%)
amplitude points of a negative pulse.
If MCross1Polarity = ‘–’
then
NegativeWidth = (MCross2MCross1)
else
NegativeWidth = (MCross3MCross2)
Waveform[Start]
See Mean dBm on page B–10.
Amplitude measurement. The absolute difference between the maximum and
minimum amplitude.
PeaktoPeak = MaxMin
Negative Duty Cycle
Negative Overshoot
Negative Width
Optical Power
Peak to Peak

Table of Contents

Other manuals for Tektronix TDS 784D

Related product manuals