Data input method User input
Location Main Acquisition board
User accessible No
To clear Remove the battery from the board.
Process to sanitize Remove the battery from the board.
USB2
Type and size USB 2.0 OTG host/peripheral controller, 8 K X 16 RAM
Function Contains USB controller program and data memory
Type of user information
stored
None
Backed up by battery? No
Method of modification Indirect
Data input method Firmware operations
Location Main Acquisition board. TBS and TDS models only.
User accessible No
To clear Remove power from the instrument for at least 20 seconds.
Process to sanitize Remove power from the instrument for at least 20 seconds.
SRAM 64 K
Type and size SRAM, 64 K X 16
Function Data memory for the USB processor
Type of user information
stored
None
Backed up by battery? No
Method of modification Indirect
Data input method Firmware operations
Location Main Acquisition board. TBS and TDS models only.
User accessible No
To clear Remove power from the instrument for at least 20 seconds.
Process to sanitize Remove power from the instrument for at least 20 seconds.
Memory devices Volatile memory devices (cont.)
2 Declassification and Security Instructions

Other manuals for Tektronix TDS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1000B Series and is the answer not in the manual?

Tektronix TDS1000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals