Press or to stop the calibration and exit the interface(fig.20 & fig.21).If the calibration
progress is 100%, the calibration is completed and press or button to exit the interface.
Calibration
CO
2
Zero
HCHO Zero
CO2 Sensor under calibration
Calibration
CO
2
Zero
HCHO Zero
HCHO Sensor under calibration
Please put the sensor outdoors for
5 minutes until its calibration is
finished.
fig.20
Calibration
CO
2
Zero
HCHO Zero
HCHO Sensor under calibration
HCHO zero point calibration
successful.
HCHO Sensor under calibration
100%
6%
100%
Calibration
CO
2
Zero
HCHO Zero
CO2 Sensor under calibration
Please put the sensor outdoors for
5 minutes until its calibration is
finished.
6%
CO2 zero point calibration
successful.
Calibration
TEMP
23.0 ℃
OFF
- 0.2 ℃
HUM 33.0 %RH
OFF
+1.5 %RH
Press the Up or Down button
to set the current value,
press the menu button to
switch to the next item.
Calibration
TEMP
25.0 ℃
OFF
+1.8 ℃
HUM 36.0 %RH
OFF
+4.5 %RH
Press the Up or Down button
to set the current value,
press the menu button to
switch to the next item.
fig.19
Formula: OFF(TEMP target value) = TEMP(target value) - TEMP(current value) + OFF(TEMP current value).
OFF(HUM target value) = HUM(target value) - HUM(current value) + OFF(HUM current value).
Example (fig.19):
If TEMP target value is 25.0, current value is 23.0, current value of OFF(TEMP) is -0.2.
Then "OFF(TEMP target value)=25.0-23.0+(-0.2)=+1.8".
If HUM target value is 36.0, current value is 33.0, current value of OFF(HUM) is +1.5.
Then "OFF(HUM target value)=36.0-33.0+(+1.5)=+4.5".
Increase "OFF(TEMP)" to +1.8°C
Increase "OFF(HUM)" to +4.5%rh
10
4. HCHO Zero Calibration:
3. CO
2 Zero Calibration:
fig.21
Note: 1. Please ensure that you can calibrate under fresh air outdoors.
2. During the calibration process, please do not perform other operations until the
calibration is successful and you return to the previous level screen.
2. TEMP & HUM Calibration: You can calibrate the PM2.5 value by adjusting the "K" factor.