What to do if the interrupted light is on for Terumo Harvest SmartPReP 2 Laboratory Equipment?
- MMichelle CarpenterAug 1, 2025
If the interrupted light is on for your Terumo Laboratory Equipment, it might be due to several reasons. First, ensure the lid is opened to clear the interrupt circuit, especially on the first power up. If the Stop button was accidentally pressed, press the Lid Open button, close the lid, and try again. Also, check if the display label is touching the Stop button; if so, cycle the Stop button and retry. In case of a power drop, press the Lid Open button, close the lid, and retry. If these steps don't resolve the issue, try cycling the power a few times; for further assistance, contact Harvest Tech Service at 877-842-7837.