Outputs
AC output
(using selected frequency weighting)
Output voltage:
2Vrms (at full-scale of the range)
Output impedance:
5k
Ω
Load impedance:
1M
Ω
DC output
Output voltage:
10mV/dB
Output impedance:
5k
Ω
Load impedance:
1M
Ω
I/O connector:
Sound level meter control from and data output to a computer (USB)
Alarm output:
5Vdc, typical
Power requirements
Qty 4 x 1.5V IEC R6P (size “AA”) manganese super heavy duty batteries or
equivalent.
Battery life
: Approx. 24 hours
External power source:
DC voltage from 5V to 6V
Current rating:
Approx. 20mA @ 6V
Ambient conditions:
Operating conditions
: -10 to +50, 30% to 90%RH non-condensing
Storage conditions
: -10 to +60, <70%RH non-condensing
Effect of temperature:
< 0.5dB (-10 to +50)
Effect of humidity:
< 0.5dB (for 30%RH to 90%RH at 40, 1000Hz)
Effect of vibration:
A 40 Hz 1m/s vibration produces no noticeable effect.
Effect of magnetic field:
No noticeable effect.
Compliance with standards:
:
indicates
compliance with applicable European Union Directives.
EMC Emission
: IEC 61000-6-3, Generic emission standard for residential,
commercial and light industrial environments.
No significant emissions from the instrument.
IEC 61672-1, Instrumentation standard classification group X and
performance class 2 sound level meter.
EMC Immunity
: IEC 61000-6-2, Generic standard-Immunity for industrial
environments.
No degradation in performance when subjected to 10V/m
unmodulated.