26
1100.. EErrrroorr mmeessssaaggeess
If problems occur which are not described here, please contact Testo or your local distributor. For
contact data, see back of this document or web page www.testo.com/service-contact
Possible ccauses
Battery power too low
Battery empty
Measuring program was
cancelled by Reset
An attempt was made to
start the data logger via
GO
: data logger indicates
that
Start criterion PC Start
is programmed.
An attempt was made to
start the data logger via
GO
: data logger indicates
that
Start criterion
Date/Time Start
is
programmed.
The
GO
button was
pressed for more than 3
seconds in the
END
mode
Remedy // CComments
Read out data and change battery
(See
9. Changing bbattery
, P. 25)
Communication with PC is possible. Communi-
cation with testo 575 fast printer/ testo 580 data
collector is not possible.
Change battery (See
9. Changing
battery
, P. 25)
No communication with PC / testo 575 fast
printer/ testo 580 data collector possible.
Reactivate data logger via testo ComSoft
software
Start the data logger via PC
or change the
Start criterion
to
Key start
.
Change
Start criterion
to
Key start
.
No function is allocated.
Error mmessage
OFF
and
END
are lit,
flashes
OFF
is lit
OFF
,
END
and
are lit
PC
is lit
date
is lit
Red and green LEDs
flash alternately five
times