Using the product8
Carrying out gas detection:
Attention! Destruction of the sensor by non-desorbant substances (e.g. oils)!
> Do not use the instrument in dirty surroundings.
> Move the sensor head as close as possible, and slowly (3 to 5 cm per second)
over the parts which are to be examined for leaks.
Resetting maximum value display:
> Press and simultaneously.
Switching instrument off:
> Press and hold
until the display switches off.