Temperature correction value table
Relative humidity: 65 %RH
Reference value sound pressure level: 124 dB
Temperature range with discrepancy <0.5 dB: 10...40 °C
Temperature in °C Correction in dB
-10 -0.7
0 -0.7
5 -0.6
50 1
Frequency weighting
Nom. frequency in Hz A-weighting C-weighting Error limits
in Hz in dB in dB Class 2 in dB
31.5 -39.4 -3.0 ±3
125 -16.1 -0.2 ±1.5
250 -8.6 -0.0 ±1.5
500 -3.2 -0.0 ±1.5
1000 0 0 ±1.5
2000 +1.2 -0.2 ±2
4000 +1.0 -0.8 ±3
8000 -1.1 -3.0 ±5
8. Accessories and Spare Parts
Name Item no.
testo 816 sound level meter 0563 8165
incl. battery, Instruction Manual, screwdriver, wind cap
Calibrator 0554 0452
Wind cap 0193 0816
9V rechargeable battery 0515 0025
Recharger for external recharging of rechargeable battery 0554 0025
Mains unit 0554 1084
Screwdriver 0554 0818
7. Technical Data