6 Using the product
31
Important settings
Before recording an image, check whether the protection glass
option is set correctly to prevent corruption of the measurement
results, see Optics... page 29.
Before saving an image, ensure that it is focused correctly (in
focus), see Manually focusing the image page 18. Images that are
not in focus cannot be corrected retroactively!
To obtain precise readings, the emissivity and reflected
temperature must be set correctly, see Set emissivity/reflected
temperature page 27. A subsequent adjustment is possible via the
PC software as required.
With an activated auto scaling, the colour scale is continuously
adapted to the min./max. values of the current measurement
image. The colour that is to be assigned to a specific temperature
thereby changes constantly. To be able to compare multiple
images by means of the colouring, the scaling must be manually
set to fixed values, see Scale... page 24, or must be subsequently
adjusted to uniform values using the PC software.
Select the storage location (folder):
1. While an image is frozen (fixed image): Press the left [Folder]
quick select button.
- The Folder dialogue is opened.
To create a new folder, see Create new folder page 23.
2. Select the desired folder with the joystick and open with [OK].
- The folder name of the opened folder is displayed in the header.
3. Press left [Apply] quick select button to select the opened
folder.