¬ Multi-purpose clip
Á Swivel head. The dis -
-play can always be
read.
Commissioning
Remove the protective film
on the display.
Switching on
Press button once:
Segmenttest
Current reading
Switch measurement view
Press button repeatedly:
Hold Current reading
Measuring in liquids:
Immerse measuring tip at
least 4,5 cm.
Switching off
Keep button pressed for 3 seconds.
Selecting between °C/°F
When switching on keep button pressed until the
segment test is finished, instrument switches to
configuration menue. Press button again the
instrument alternates between °C and °F. Wait
2 sec. before pressing the button again. The
selected unit then applies.
Technical data
Type of application: .................................. Short measurement
Sensor: ..................................................Thermocouple, type K
Measurement range: ...... -50 to +350°C; short-term to +500°C/
-58 to +660°F; short-term to +930°F
Parameters: .....................................................................°C/°F
Resolution: ............................................................0.1°C/0.1°F
Accuracy: ...... (referred to calibration temperature of 25°C)
............................±1 °C (-50...99,9 °C)/±1.8°F (-58 to +212°F)
.......................................................±1% of m.v. (rest of range)
Ambient temperature: ..................0 to +40°C/+32°F to +104°F
Storage temperature: .................... -20 to +70°C/-4 to +160°F
Response time t99: .............................. Approx. 10 s (in water)
Battery type: ................................................................3 x AAA
Battery lifetime: ................................................Approx. 1000 h
Probe stem: ..........................................Ø 0.1 in / length: 7.9 in
CE guideline ......................................................... 2014/30/EC
Please read before using instrument
Do not measure on live parts!
Risk of injury from measuring tip
Observe measurement ranges of sensor
Do not exceed maximum storage and operating
temperature (e.g. protect measuring instrument from
direct sunlight)
Approx. 15 min. adjustment time if ambient
temperature changes
Inexpert handling cancels your warranty.
Changing the battery
The battery has to be changed if the symbol
appears in the display during the
measurement.
3 Instrument is switched off.
1 Open battery compartment cover.
2 Remove spent batteries and insert new batteries
(3 x AAA) into the battery compartment. Observe
polarity!
3 Close battery compartment cover.
Auto Off function
If no button has been pressed in approx. 10
minutes, the instrument switches itself off
automatically.
Instruction manual en
Penetration Thermometer · testo 905-T1
ø 0.1 in
Excessively hot measuring
medium >350 °C!
Damage to the housing!
Please note maximum 180
mm immersion depth of the
probe shaft!
Attention!
7.9 in