XT1000 Quick Start Guide
Page 2 of 8
Press on dEF and you will then see CAP, which is the maximum
capacity of your scale that you will require.
Press again and this allows you to input the maximum capacity of
your scale using up to 6 digits. You input this number using the ,
and keys.
Once you have entered your maximum capacity press to store this
value and you should again see CAP.
Note: CAP ÷ D1 ≤ 100000 or you will see ErrCAP.
Press to move to d1, which is the division of the scale that you
require. This division is determined by taking the higher value of:
i. CAP (value you have used for the maximum capacity of your
scale) ÷ 100,000
ii. (Load cell capacity, expressed in the same unit of measure as used for CAP
÷ 10,000) x √(number of load cells)
iii. Maximum product load ÷ division of the load cell (e.g. 3000
division for a C3 cell.)
Once you have calculated the highest value, press and using the
or keys choose the next closest value (rounding up) from
1, 2, 5, 10, 20 or 50. If you require a number smaller than 1 you can
choose the DP (decimal place) in the next step, for example if 0. 05 is
required choose 5 in d1 and then 0. 00 in dP.
Once you have chosen your division press to store and you will then
again see d1.
Note: CAP ÷ D1 ≤ 100000 or you will see Errd1