EasyManua.ls Logo

Thames Side XT2000 - User Manual

Thames Side XT2000
8 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
XT2000 Quick Start Guide
Page 1 of 8
Turn your XT2000 on and wait until the instrument has fully completed
its countdown sequence and the display has stabilised.
If turning on for the first time it is advisable to leave the XT2000 for 30
minutes prior to calibration.
Press both the and keys together which will change the display
to Id0000 with the first 0 flashing.
Pressing the up key ( ), allows you to change each 0 to 1, 2, 3 etc.
Pressing the left ( ) and right ( ) keys then moves you to
each position. E.g. _000, 2_00, 28_0, 280_
Change these 0’s to read 2802 which is the factory access code.
Once you have changed Id0000 to read Id2802 press the enter
button.
The XT2000 will then briefly display UnPro to confirm that the
instrument is in the unprotected mode and the parameters can be
altered.
Once it has shown UnPro the display will change to show to dEF (Scale
Definition)
In order to calibrate your system you will first have to define the scale.
For more detail please see section 3.2 in the user manual.
Question and Answer IconNeed help?

Do you have a question about the Thames Side XT2000 and is the answer not in the manual?

Thames Side XT2000 Specifications

General IconGeneral
BrandThames Side
ModelXT2000
CategoryMeasuring Instruments
LanguageEnglish

Summary

Scale Configuration

Entering Factory Access Code

Input '2802' to access calibration and configuration settings.

Setting Maximum Capacity (CAP)

Define the scale's maximum allowable weight capacity.

Setting Scale Division (d I)

Configure the smallest measurement unit for the scale.

Setting Decimal Place (dP)

Specify the number of decimal places for readings.

Configuring Auto-Zero Band (0-ErAC)

Define the range for automatic zeroing based on weight stability.

Setting Zero Range (0-top)

Define the acceptable range for the scale's zero point.

Enabling Auto-Zero Function (0.5ŁArt)

Enable or disable the indicator's auto-zero feature.

Setting Under Limit (UnderL)

Define the minimum input threshold for error signals.

Selecting Unit of Weight (Unit)

Choose the desired unit of measurement for the scale.

Calibration with Verified Weights (CAL 1)

Initiate Verified Calibration (CAL 1)

Select CAL 1 mode for calibration using verified weights.

Set Zero Point for CAL 1

Establish the scale's zero point before applying weights.

Apply Verified Weight

Place a known weight on the scale for calibration.

Confirm CAL 1 Reading

Verify the scale reading matches the applied weight.

Theoretical Calibration (CAL 2)

Initiate Theoretical Calibration (CAL 2)

Select CAL 2 for calibration without verified weights.

Input Load Cell Capacity (LCAP)

Enter the nominal capacity of the load cell(s).

Input Number of Supports (Lno)

Enter the total count of load supports, including empty ones.

Input mV;v Value (LSn)

Enter the mV/v value of the load cell(s).

Perform Zero Adjustment

Execute the automatic zero adjustment after calibration.

Related product manuals