Press to see LSn which is the mv/v value of the load cell(s) being
used. Typically this information is on a sticker on the load cell.
If you have more than one cell in your system then average out the
mv/v values and use this average figure.
Press to enter the mv/v value using the , and keys,
Once correct press to store this value and you will return to LSn.
Press to see 0Ero which is your automatic Zero adjustment.
Press the and you will see the preset coefficient value.
Press again you should see -CALI- flashing which will store this
as your zero point.
Normally this is sufficient for most calibration purposes and at this point
you can press and again to leave the configuration level
completely and your X2000 should now be calibrated and ready to
weigh.