THEORETICAL CALIBRATION (CAL 2) when no verified weights are available.
Section 4.2 in the user manual
Ensure that there is no unwanted weight on the system and that you
are in the configuration level.
Using the or keys scroll from dEF until you see CAL 2.
Press and you will see LCAP.
Press and enter the nominal capacity of one of the load cells being
used.
This value is expressed in the same decimal point used in CAP and d1
Press to store this value and return to LCAP.
Press to see Lno, which is the number of load supports.
Please note that all supports must be counted including those without
load cells.
Press to enter the number of supports using the button.
When correct press to store this value and you will return to Lno.