Thames Side XT2000 - Setting Under Limit (UnderL); Selecting Unit of Weight (Unit)

Thames Side XT2000
8 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XT2000 Quick Start Guide
Page 4 of 8
Press to see UndErL, the point at which the computer indicates
the input error signal below the minimum range.
Press and choose your option using the or keys, the
default is setting is oUErL, once you have chosen your option press
to confirm and go back to UndErL.
Press to see UnIt, press to choose your unit of weight using
the and keys. The options are:
K 9 = Kilogram,
t = Ton,
g = Gram,
Lb = Pound,
o = Ounce,
nonE = None)
Press to confirm your choice and go back to UnIt.
The XT2000 scale has now been defined.
By pressing you will return to dEF where pressing either the or key and scroll until you see:
CAL 1, to perform a calibration with verified weights. (Section 4.1 in the user manual)
CAL 2, to perform a theoretical calibration if you do not have any verified weights. (Section 4.2 in the user manual)
Pressing again when you are on dEF exits the configuration level completely.
Note: If you exit the configuration level you will have to re-enter the Factory Access Code 2802 to calibrate your XT2000.