ELECTRONIC PERSONAL DOSEMETER HANDBOOK
EPD/HB/40521/000
APPENDIX B
ISSUE 5
Thermo
Page 93
APPENDIX B
ELECTROMAGNETIC COMPATIBILITY
Standard Test
The EPD was tested to the following specification with no failures:
Test Type Specification Test Frequency
Range
Level Notes
Radiated
Susceptibility E-
Field
Mil Std 461D RS103 10kHz – 40 GHz 200V/m
Radiated
Susceptibility E-
Field
Def Stan 59-41 DRS02 14kHz – 18GHz 200V/m
Radiated
Susceptibility H-
Field
Mil 461D RS101 30Hz – 100KHz See Chart 1 Note 1
Electrostatic
Discharge
EN61000-4-2 ESD N/A 8KV
Radiated Electric
Field Emissions
EN 50081-1 N/A 30MHz – 1 Ghz 30dB below
Limit
Radiated
Susceptibility E-
Field High Field
Spot
Frequencies CW
N/A 100kHz–40 Ghz See Chart 2 Note 2
Radiated
Susceptibility E-
Field High Field
Spot
Frequencies
CW with 1kHz 1
µS square wave
N/A 500MHz–40 Ghz See Chart 3 Note 2
Notes
:
1. The limits are as Mil461D (Army) between 30Hz and 1kHz. Between 1kHz and
100kHz the limit is increased by 20 dB.
2. The units were tested up to the point of failure or to the maximum field
attainable in the test configuration.
Graphs follow: