ELECTRONIC PERSONAL DOSEMETER HANDBOOK
EPD/HB/40521/000
OPERATING INSTRUCTIONS
ISSUE 5
Thermo
Page 19
Priority
Alarm/Event
Alarm Tone
Frequency
Volume
1 EPD Failure Alarm. Intermittent Single Slow Tone
__ __ __ __
High Loud
2 Over-range Alarm. Continuous Dual Fast Tone
_-_-_-_-_-_
High/Low Loud
3 Hp (10) Dose Alarm
(2)
Continuous Single Tone
____________
High Loud
4 Hp (0.07) Dose
Alarm
Continuous Single Tone
____________
High Loud
5 Hp (10) Dose Alarm
(1)
Continuous Single Tone
____________
High Loud
6 Hp (10)/h Dose
Rate Alarm (2)
Continuous Single Tone
____________
High Loud
7 Hp (0.07)/h Dose
Rate Alarm
Intermittent Single Fast Tone
_ _ _ _ _ _
High Loud
8 Hp (10)/h Dose
Rate Alarm (1)
Intermittent Single Fast Tone
_ _ _ _ _ _
High Loud
9 Battery Alarm Intermittent Single Slow Tone
__ __ __ __
Low Quiet
10 Return Alarm Intermittent Single Slow Tone
__ __ __ __
Low Quiet
11 Count Down Alarm Continuous Single Tone
____________
High Loud
12 Abuse Alarm Intermittent Single Slow Tone
__ __ __ __
Low Quiet
Table 3.3 - Default Alarm Tones