2007.03.07Flow2 stairliftTab 06: Systematic fault finding
www.ThyssenKruppAccessibility.nl20
Error messages
Possible
solution
- Switch the lift off and on.
- Reprogram the safety processor.
- Replace the main board if none of the above solutions has helped or if the
fault keeps coming back. This is an extreme measure.
Reset Switch off and on.
Modes 0, 2
Error code E06
Type of error System error
Error message Safety mcu programming failure
Description Programming the safety processor failed.
Priority Highest
Possible
solution
- Switch the lift off and on.
- Reprogram the safety processor.
- Replace the memory card.
- Replace the main board if none of the above solutions has helped or if the
fault keeps coming back. This is an extreme measure.
Reset Switch off and on.
Modes 0, 2
Error code E10
Type of error System error
Error message Relay circuit error
Description This error indicates that the main power relay on the power board does not
switch on and off correctly. This can be the result of sticking contacts or
defective parts.
Priority Average
Possible
solution
- Replace the power board.
- If the error persists, replace the cassette.
Reset Switch off and on.
Modes 0, 1, 2, 3, 4, 5, 6