G1/G2 enable/disable group execute trigger
G1 - Enables Group Execute Trigger response.
G2 - Disables Group Execute Trigger response.
Group Execute Trigger (GET) is a command issued by the bus controller to make sev-
eral devices respond simultaneously. The devices having been previously told how to
respond when the GET command is received.
EXAMPLE (HP BASIC)
10 REMOTE 708 !OPEN IEEE ADDRESS 8
20 OUTPUT 708;”G1” !SET 9814 TO RESPOND TO GET
30 OUTPUT 708;”R2” !COMMAND TO EXECUTE
40 TRIGGER 708 !EXECUTE COMMAND
50 OUTPUT 708;”G2” !COMMAND TO EXECUTE
60 TRIGGER 708 !EXECUTE COMMAND TO EXIT GET MODE
K1/K2 enable/disable front panel control
K1 - Set the 9814 to respond to front panel control
K2 - Disables front panel controls
P (n) set percentage deviation
Set deviation on present output in range +9.999 and –9.999.
EXAMPLE (CBM BASIC)
10 PRINT #1,”P3.456” REM SET output 3.456% high
20 PRINT #1,”P-0.02” REM SET output 0.02% low
R1 to R4 range setting commands
Selects a voltage range from IEEE command table.
COMMAND
RANGE PROGRAM
SELECTED IN
R1 10mV mVolts
R2 100Mv mVolts
R3 1V Volts
R4 10V Volts
RA: Automatically selects the range for the output required. Default : R1
T1/T2 select transmission terminator
Sets the last character appended to a transmission from the 9814