Calibration Mode
CAL
RESULTRESULT
+ x + x
CAL
4
User calibration
Note
:
user calibration (Display in calibration mode)
factory calibration (Display in calibration mode)
After confirming the selection with the [MODE] key the instrument will
show CAL/LR.
Scroll through ranges using the [MODE] key.
Place the blank in the adapter making sure that the marks
are aligned.
Press the [ZERO/TEST] key.
The "Range" symbol flashes for approx. 8 seconds.
The display shows the following in alternating mode:
Perform measurement with a standard of known concentration as
described.
Press the [ZERO/TEST] key.
The "Range" symbol flashes for approx. 3 seconds.
The result is shown in the display, alternating with CAL.
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
By pressing the [ON/OFF] key, the new correction factor is calculated
and stored in the user calibration software.
Confirmation of calibration (3 seconds).
Mode
On
Off
Zero
Test
: :
Note: The instrument calibration in the range MR is automatically taken also for HR.
RESULTRESULT
CAL
LR
0.0.0
CAL
CAL
cAL
Zero
Test
Zero
Test
Time Cal
Store Date
4
Cal
RANGERANGE
RANGERANGE
!
Proceed to the next value to be edited by pressing [!] key.
After setting the minutes and pressing the [!] key the display will show
“IS SET” and the instrument returns to the measurement mode.