Technical Data
Technical Data
Instrument double wavelength, automatic wavelength selection,
direct reading colorimeter
Light source: LEDs, interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
430 nm
= 5 nm
610 nm
= 6 nm
Wavelength accuracy ± 1 nm
Photometric accuracy* 3% FS (T = 20° C – 25° C)
Photometric resolution 0.01 A
Power supply 4 batteries (AAA/LR 03)
Operating time 17hr operating time or 5000 measurements in constant
use when display backlight and Bluetooth
®
data transfer
are off.
Auto-OFF automatic switch off
10 minutes after last keypress
Display backlit LCD (on keypress)
Storage internal ring memory for 125 data sets
Serial Interface Bluetooth
®
4.0 for the transfer of stored measurement data
Bluetooth
®
Modul: Specifications:
Modul: BLE113-A
Bluetooth
®
4.0 LE
FCC ID: QOQBT113
IC: 5123A-BGTBLE113
Time real time clock und date
Calibration user and factory calibration
resetting to factory calibration possible
Dimensions 155 x 75 x 35 mm (LxWxH)
Weight approx. 260 g (incl. batteries)
Ambient conditions temperature: 540°C
rel. humidity: 3090% (non-condensing)
Waterproof floating; as defined in IP 68 (1 hour at 0.1 meter)
CE Certificate for Declaration of CE-Conformity
at www.lovibond.com
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.