4 VIEWING CONDITIONS CHECK
4.1 Objective
To keep and maintain the optimal viewing conditions on Monochrome Monitors for
Mammography
4.2 Frequency
Daily
4.3 Required Equipment
Monochrome Monitors for Mammography, display monitor
Serial cable for communicating between PC and display monitor (for ME511L)
USB cable for communicating between PC and display monitor (for other models)
4.4 Procedure
1. Review the data form “Form 3: Viewing Conditions Check and Setting” on P15
(completed by the Medical Physicist during the most recent evaluation) posted in
the reading room. Refer this form as optimal during your future evaluation.
2. Match the room conguration where each display monitor is located to the one
specied in the data form.
3. Adjust the room conguration if the current reading room conguration differs
from the one established as optimal in the data form. (For example, if desk lights
were specied to be turned off, turn them off, or, if curtains were specied to be
closed, close them.)
4. Complete the record of the check on the form 1 in this manual.