9. SETTING UP THE UNIT:
9.1 Before you begin using the unit, it is important to define your measuring
requirements. These will allow the unit to display accurately on the
condition of sample that is being measured.
9.2 After the unit is switched ON, press and hold down the key until
display shows tr
trtr
tr.
9.3 Press to go to next parameter and for previous parameter.
Below are the description of each parameter setting:
(Anytime during setup, press to cancel or return to measuring mode)
tr
(P1
P1P1
P1.0
00
0) DATA TRANSMISSION TO PERSONAL COMPUTER
This setting allows you to download saved memory data to PC via the
RS232 connection.
1. Download a free software from the internet at the below link:
http://www.eltima.com/download/rs232-data-logger
2. Connect the RS232 cable to the back of the meter and PC RS232
socket. Optional USB interface connector can be used.
3. Install and RUN the application on the PC.
4. Under Available ports, select the com port meter is connected to.
5. Click to select file location and type file name then click save.
6. Set Baud-rate to 9600, Data-bits to 8, Parity to non, Stop-bits to 1,
Flow control to non and Click
7. Press on the meter and display will show
out
outout
out
(P1
P1P1
P1.1
11
1)
flashing, PC will indicate transmission in progress.
8. When transmission is completed, display will return to tr P1
tr P1tr P1
tr P1.0
00
0.
9. On the PC, click and a text file will be created.
10. Locate the directory and open the file in text format. Data is
display with time and date for analysis.
11. Press to go to next parameter.
Clr
(P2
P2P2
P2.0
00
0) CLEAR MEMORY
This setting will clear all the saved memory in the unit.
1. Press and display will show n
nn
no
o o
o (P2
P2P2
P2.1
11
1) flashing.
2. Press once to select between YES
YESYES
YES or no
nono
no. If you select YES
YESYES
YES
and press , all memory will be cleared. Otherwise select no
nono
no
then press , display will return to CLr
CLrCLr
CLr (P2
P2P2
P2.0
00
0).
3. Press to go to next parameter.
. . .